Reducing SAR and enhancing cerebral signal-to-noise ratio with high permittivity padding at 3 T

Magn Reson Med. 2011 Feb;65(2):358-62. doi: 10.1002/mrm.22695. Epub 2010 Nov 30.

Abstract

Previous works have shown that placement of a high-dielectric pad can improve image intensity in a region adjacent to the pad, or that placement of dielectric pads around a large surface of the head can improve image homogeneity on an entire plane through the head in high-field MRI. Here, experimental results show that use of high-dielectric pads around the human head can reduce the required input radiofrequency power by 50% while enhancing image signal-to-noise ratio by 20-40% throughout the cerebrum at 3 T. Thus, dielectric pads may be used to provide a relatively simple and low-cost method for improving quality and safety of MRI in a variety of applications at 3 T.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain / anatomy & histology*
  • Computer Simulation
  • Humans
  • Image Enhancement / instrumentation
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Radio Waves