Determination of the depth dose distribution of proton beam using PRESAGE™ dosimeter

J Phys Conf Ser. 2010;250(1):12035. doi: 10.1088/1742-6596/250/1/012035.

Abstract

PRESAGE™ dosimeter dosimeter has been proved useful for 3D dosimetry in conventional photon therapy and IMRT [1-5]. Our objective is to examine the use of PRESAGE™ dosimeter for verification of depth dose distribution in proton beam therapy. Three PRESAGE™ samples were irradiated with a 79 MeV un-modulated proton beam. Percent depth dose profile measured from the PRESAGE™ dosimeter is compared with data obtained in a water phantom using a parallel plate Advanced Markus chamber. The Bragg-peak position determined from the PRESAGE™ is within 2 mm compared to measurements in water. PRESAGE™ shows a highly linear response to proton dose. However, PRESAGE™ also reveals an underdosage around the Bragg peak position due to LET effects. Depth scaling factor and quenching correction factor need further investigation. Our initial result shows that PRESAGE™ has promising dosimetric characteristics that could be suitable for proton beam dosimetry.