Purpose: The aim of the present study was to compare intensity-modulated radiotherapy (IMRT) with volumetric arc therapy (VMAT), in the treatment of prostate cancer with maximal dose escalation to the intraprostatic lesion (IPL), without violating the organ-at-risk constraints. Additionally, the use of 6-MV photons was compared with 18-MV photons for all techniques.
Methods and materials: A total of 12 consecutive prostate cancer patients with an IPL on magnetic resonance imaging were selected for the present study. Plans were made for three IMRT field setups (three, five, and seven fields) and one VMAT field setup (single arc). First, optimal plans were created for every technique using biologic and physical planning aims. Next, an additional escalation to the IPL was planned as high as possible without violating the planning aims of the first step.
Results: No interaction between the technique and photon energy (p=.928) occurred. No differences were found between the 6- and 18-MV photon beams, except for a reduction in the number of monitor units needed for 18 MV (p<.05). All techniques, except for three-field IMRT, allowed for dose escalation to a median dose of ≥93±6 Gy (mean±standard deviation) to the IPL. VMAT was superior to IMRT for rectal volumes receiving 20-50 Gy (p<.05).
Conclusion: VMAT allowed for dose escalation to the IPL with better sparing of the rectum than static three-, five-, and seven-field IMRT setups. High-energy photons had no advantage over low-energy photons.