Coordinated basal-bolus infusion for tighter postprandial glucose control in insulin pump therapy

J Diabetes Sci Technol. 2009 Jan;3(1):89-97. doi: 10.1177/193229680900300110.

Abstract

Background: Basal and bolus insulin determination in intensive insulin therapy for type 1 diabetes mellitus (T1DM) are currently considered independently of each other. A new strategy that coordinates basal and bolus insulin infusion to cope with postprandial glycemia in pump therapy is proposed. Superior performance of this new strategy is demonstrated through a formal analysis of attainable performances in an in silico study.

Methods: The set inversion via interval analysis algorithm has been applied to obtain the feasible set of basal and bolus doses that, for a given meal, mathematically guarantee a postprandial response fulfilling the International Diabetes Federation (IDF) guidelines (i.e., no hypoglycemia and 2 h postprandial glucose below 140 mg/dl). Hypoglycemia has been defined as a glucose value below 70 mg/dl. A 5 h time horizon has been considered for a 70 kg in silico T1DM subject consuming meals in the range of 30 to 80 g of carbohydrates.

Results: The computed feasible sets demonstrate that current separated basal/bolus strategy dramatically limits the attainable performance. For a nominal basal of 0.8 IU/h leading to a basal glucose of approximately 100 mg/dl, IDF guidelines cannot be fulfilled for meals greater than 50 g of carbohydrates, independent of the bolus insulin computed. However, coordinating the basal and bolus insulin delivery can achieve this. A decrement of basal insulin during the postprandial period is required together with an increase in bolus insulin, in appropriate percentages, which is meal dependent. After 3 h, basal insulin can be restored to its nominal value.

Conclusions: The new strategy meets IDF guidelines in a typical day, contrary to the standard basal/bolus strategy, yielding a mean 2 h postprandial glucose reduction of 36.4 mg/dl without late hypoglycemia. The application of interval analysis for the computation of feasible sets is demonstrated to be a powerful tool for the analysis of attainable performance in glucose control.

Keywords: glucose control; insulin pump therapy; interval analysis; set inversion; type 1 diabetes mellitus.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Glucose / drug effects*
  • Blood Glucose / metabolism
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Diabetes Mellitus, Type 1 / metabolism
  • Humans
  • Hypoglycemic Agents / administration & dosage*
  • Hypoglycemic Agents / metabolism
  • Insulin / administration & dosage*
  • Insulin / metabolism
  • Insulin Infusion Systems*
  • Models, Theoretical*

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin