Size of the proximal neck in AAAs treated with balloon-expandable stent-grafts: CTA findings in mid- to long-term follow-up

J Endovasc Ther. 2009 Dec;16(6):696-707. doi: 10.1583/09-2711.1.

Abstract

Purpose: To determine the evolution of the proximal aortic neck diameter in mid- to long-term follow-up after endovascular aneurysm repair of abdominal aortic aneurysm (AAA) with a balloon-expandable stent-graft.

Methods: Thirty patients (27 men; average age 71 years, range 56-87) with infrarenal AAAs were treated with the SETA-Latecba balloon-expandable stent-graft (6 aortomonoiliac and 24 bifurcated configurations). Follow-up ranged from 4 to 8 years (mean 73.4 months). Computed tomography was done systematically before the procedure, after implantation (1-3 months), at 1 year, and annually thereafter. The last follow-up scan was utilized to measure the proximal neck for purposes of comparison with baseline and the initial post-implant scans.

Results: Five patients died during follow-up of causes unrelated to the procedure. No endoleaks or graft migrations were observed. The pre-deployment proximal neck diameter (a) averaged 23.4 mm (range 18-32), the diameter after deployment of the stent-graft (b) averaged 24.9 mm (range 18-34), and the most recent follow-up proximal neck measurement (c) averaged 23.8 mm (range 18-31). Comparing the last follow-up to the post-implant measurements (c-b), the neck diameter decreased in 15 (50%) patients [7 with short necks (i.e., <15 mm)] and remained unchanged (no variation) in 15 (50%) patients (4 with short necks). All patients treated with the SETA-Latecba balloon-expandable stent-graft showed stability of the proximal aortic neck diameter in mid- to long-term follow-up.

Conclusion: The study showed that the diameter reached at initial deployment did not increase further in the long term, which supports the safety and reliability of this modular balloon-expandable stent-graft and illustrates that this device does not produce dilatation of the proximal neck after deployment. Future dilatation of the aortic neck is unlikely, and consequently, migration or delayed type I endoleak are also unlikely.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal / diagnostic imaging*
  • Aortic Aneurysm, Abdominal / surgery*
  • Aortography / methods*
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Catheterization*
  • Female
  • Foreign-Body Migration / etiology
  • Foreign-Body Migration / prevention & control
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Failure
  • Retrospective Studies
  • Stents*
  • Time Factors
  • Tomography, X-Ray Computed*
  • Treatment Outcome