An artificial gastrocnemius for a transtibial prosthesis

Annu Int Conf IEEE Eng Med Biol Soc. 2009:2009:5034-7. doi: 10.1109/IEMBS.2009.5333697.

Abstract

A transtibial amputee does not have a functional gastrocnemius muscle, which affects the knee as well as the ankle joint. In this investigation, we developed a transtibial prosthesis comprising an artificial gastrocnemius mechanism as well as a powered ankle-foot device. A pilot study was conducted with a bilateral transtibial amputee walking at a self-selected speed. The trial compared muscle electromyography and metabolic cost data for the amputee while using the active gastrocnemius prosthesis and a conventional Flex-Foot prosthesis. The experimental data showed that the compensation for ankle-foot and gastrocnemius function offered by the active device resulted in a reduced metabolic cost for the amputee participant.

MeSH terms

  • Amputees / rehabilitation*
  • Ankle Joint / physiology
  • Artificial Limbs*
  • Foot / physiology
  • Humans
  • Prosthesis Design
  • Tibia / surgery*