The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study

Circ Arrhythm Electrophysiol. 2009 Oct;2(5):474-80. doi: 10.1161/CIRCEP.109.849638. Epub 2009 Aug 4.

Abstract

Background: It is unknown if brief episodes of device-detected atrial fibrillation (AF) increase thromboembolic event (TE) risk.

Methods and results: TRENDS was a prospective, observational study enrolling patients with > or = 1 stroke risk factor (heart failure, hypertension, age > or = 65 years, diabetes, or prior TE) receiving pacemakers or defibrillators that monitor atrial tachycardia (AT)/AF burden (defined as the longest total AT/AF duration on any given day during the prior 30-day period). This time-varying exposure was updated daily during follow-up and related to TE risk. Annualized TE rates were determined according to AT/AF burden subsets: zero, low (<5.5 hours [median duration of subsets with nonzero burden]), and high (> or = 5.5 hours). A multivariate Cox model provided hazard ratios including terms for stroke risk factors and time-varying AT/AF burden and antithrombotic therapy. Patients (n=2486) had at least 30 days of device data for analysis. During a mean follow-up of 1.4 years, annualized TE risk (including transient ischemic attacks) was 1.1% for zero, 1.1% for low, and 2.4% for high burden subsets of 30-day windows. Compared with zero burden, adjusted hazard ratios (95% CIs) in the low and high burden subsets were 0.98 (0.34 to 2.82, P=0.97) and 2.20 (0.96 to 5.05, P=0.06), respectively.

Conclusions: The TE rate was low compared with patients with traditional AF with similar risk profiles. The data suggest that TE risk is a quantitative function of AT/AF burden. AT/AF burden > or = 5.5 hours on any of 30 prior days appeared to double TE risk. Additional studies are needed to more precisely investigate the relationship between stroke risk and AT/AF burden.

Trial registration: ClinicalTrials.gov NCT00279981.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Analysis of Variance
  • Atrial Fibrillation / complications*
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / therapy
  • Chi-Square Distribution
  • Defibrillators, Implantable
  • Female
  • Humans
  • Intracranial Embolism / epidemiology
  • Intracranial Embolism / etiology
  • Ischemic Attack, Transient / epidemiology
  • Ischemic Attack, Transient / etiology
  • Male
  • Prognosis
  • Proportional Hazards Models
  • Prospective Studies
  • Risk Factors
  • Stroke / epidemiology
  • Stroke / etiology*
  • Thrombolytic Therapy

Associated data