Toward submillimeter accuracy in the management of intrafraction motion: the integration of real-time internal position monitoring and multileaf collimator target tracking

Int J Radiat Oncol Biol Phys. 2009 Jun 1;74(2):575-82. doi: 10.1016/j.ijrobp.2008.12.057. Epub 2009 Mar 26.

Abstract

Purpose: We report on an integrated system for real-time adaptive radiation delivery to moving tumors. The system combines two promising technologies-three-dimensional internal position monitoring using implanted electromagnetically excitable transponders and corresponding real-time beam adaptation using a dynamic multileaf collimator (DMLC).

Methods and materials: In a multi-institutional academic and industrial collaboration, a research version of the Calypso position monitoring system was integrated with a DMLC-based four-dimensional intensity-modulated radiotherapy delivery system using a Varian 120-leaf multileaf collimator (MLC). Two important determinants of system performance-latency (i.e., elapsed time between target motion and MLC response) and geometric accuracy-were investigated. Latency was quantified by acquiring continuous megavoltage X-ray images of a moving phantom (with embedded transponders) that was tracked in real time by a circular MLC field. The latency value was input into a motion prediction algorithm within the DMLC tracking system. Geometric accuracy was calculated as the root-mean-square positional error between the target and the centroid of the MLC aperture for patient-derived three-dimensional motion trajectories comprising two lung tumor traces and one prostate trace.

Results: System latency was determined to be approximately 220 milliseconds. Tracking accuracy was observed to be sub-2 mm for the respiratory motion traces and sub-1 mm for prostate motion.

Conclusion: We have developed and characterized a research version of a novel four-dimensional delivery system that integrates nonionizing radiation-based internal position monitoring and accurate real-time DMLC-based beam adaptation. This system represents a significant step toward achieving the eventual goal of geometrically ideal dose delivery to moving tumors.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Computer Systems
  • Feasibility Studies
  • Humans
  • Imaging, Three-Dimensional / methods*
  • Lung Neoplasms / radiotherapy
  • Male
  • Movement*
  • Particle Accelerators / instrumentation
  • Phantoms, Imaging*
  • Prostatic Neoplasms / radiotherapy
  • Prostheses and Implants
  • Radiotherapy, Intensity-Modulated / instrumentation
  • Radiotherapy, Intensity-Modulated / methods*
  • Respiration
  • Technology, Radiologic / instrumentation
  • Technology, Radiologic / methods