Evaluation of alignment error due to a speed artifact in stereotactic ultrasound image guidance

Phys Med Biol. 2008 Dec 7;53(23):N437-45. doi: 10.1088/0031-9155/53/23/N03. Epub 2008 Nov 12.

Abstract

Ultrasound (US) image guidance systems used in radiotherapy are typically calibrated for soft tissue applications, thus introducing errors in depth-from-transducer representation when used in media with a different speed of sound propagation (e.g. fat). This error is commonly referred to as the speed artifact. In this study we utilized a standard US phantom to demonstrate the existence of the speed artifact when using a commercial US image guidance system to image through layers of simulated body fat, and we compared the results with calculated/predicted values. A general purpose US phantom (speed of sound (SOS) = 1540 m s(-1)) was imaged on a multi-slice CT scanner at a 0.625 mm slice thickness and 0.5 mm x 0.5 mm axial pixel size. Target-simulating wires inside the phantom were contoured and later transferred to the US guidance system. Layers of various thickness (1-8 cm) of commercially manufactured fat-simulating material (SOS = 1435 m s(-1)) were placed on top of the phantom to study the depth-related alignment error. In order to demonstrate that the speed artifact is not caused by adding additional layers on top of the phantom, we repeated these measurements in an identical setup using commercially manufactured tissue-simulating material (SOS = 1540 m s(-1)) for the top layers. For the fat-simulating material used in this study, we observed the magnitude of the depth-related alignment errors resulting from the speed artifact to be 0.7 mm cm(-1) of fat imaged through. The measured alignment errors caused by the speed artifact agreed with the calculated values within one standard deviation for all of the different thicknesses of fat-simulating material studied here. We demonstrated the depth-related alignment error due to the speed artifact when using US image guidance for radiation treatment alignment and note that the presence of fat causes the target to be aliased to a depth greater than it actually is. For typical US guidance systems in use today, this will lead to delivery of the high dose region at a position slightly posterior to the intended region for a supine patient. When possible, care should be taken to avoid imaging through a thick layer of fat for larger patients in US alignments or, if unavoidable, the spatial inaccuracies introduced by the artifact should be considered by the physician during the formulation of the treatment plan.

MeSH terms

  • Adipose Tissue / diagnostic imaging*
  • Artifacts*
  • Calibration
  • Humans
  • Phantoms, Imaging
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Stereotaxic Techniques
  • Tomography, X-Ray Computed
  • Ultrasonography*