Parameter selection and programming recommendations for the ACE and CIS speech-processing strategies in the Nucleus 24 cochlear implant system

Cochlear Implants Int. 2002 Sep;3(2):104-25. doi: 10.1179/cim.2002.3.2.104.

Abstract

This study evaluated a range of programming parameters available for the ACE and CIS speech-processing strategies in the Nucleus 24 cochlear implant system. Specifically, the effect on speech perception of adjustments to the number of channels and the stimulation rate in the CIS strategy, and the effect of adjustments to the number of maxima in the ACE strategy were investigated in a group of adult subjects. Based on these findings, and the results of a previous study (Vandali et al., 2000), a number of recommendations for programming were identified that could improve efficiency in the clinical setting where time may be limited. The results suggest that speech perception benefits may be maximized for an individual cochlear implant recipient through concentration on selection of the appropriate stimulation rate. When using the CIS strategy, the number of channels should also be optimized. Adjustment to the number of maxima in the ACE strategy was found to be less likely to provide improvements in speech recognition for a given individual when the number of maxima parameter is set to eight or 12.