Pilot study for the standardization of insulin immunoassays with isotope dilution liquid chromatography/tandem mass spectrometry

Clin Chem. 2007 Aug;53(8):1462-9. doi: 10.1373/clinchem.2007.088393. Epub 2007 Jun 22.

Abstract

Background: An international working group convened by the American Diabetes Association (ADA) called for a reference measurement procedure for use in a trueness-based standardization project of insulin immunoassays. In view of this demand, we conducted a pilot study to investigate the feasibility of such a standardization project with our isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LC/tandem MS) procedure.

Methods: We evaluated the precision, accuracy, and limit of quantification (LoQ) of the ID-LC/tandem MS procedure by use of insulin-free serum supplemented with insulin to give 3 pools with concentrations of 0.0796, 0.769, and 5.56 microg/L. We conducted a pilot method comparison study with 4 immunoassays and 80 samples from fasting and glucose-stimulated patients.

Results: The within-run and total imprecision (CV) ranged from 3.2% to 6.3% and from 4.9% to 12.1% (listing sequence from the high to the low pool). The recovery from supplemented insulin-free sera ranged from 101.8% to 104.1%, and the LoQ was 0.07 microg/L (12 pmol/L). Weighted Deming regression and correlation analysis of the method-comparison data showed considerable between-assay variation for the immunoassays but, with the exception of one assay, excellent correlation with ID-LC/tandem MS. Recalibration of the immunoassay results considerably reduced the between-assay variation. Moreover, after recalibration, 3 of the 4 assays fulfilled the total error specification of 32% proposed by the ADA Workgroup.

Conclusions: Recalibration of insulin assays by regression equations established from method comparison with ID-LC/tandem MS can result in successful standardization and fulfillment of the total error criterion proposed by the ADA Workgroup.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Chromatography, Liquid
  • Fasting
  • Feasibility Studies
  • Glucose
  • Glucose Intolerance / diagnosis
  • Humans
  • Immunoassay / standards
  • Indicator Dilution Techniques
  • Insulin / blood*
  • Insulin Resistance
  • Obesity / diagnosis
  • Pilot Projects
  • Regression Analysis
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Mass, Electrospray Ionization
  • Tandem Mass Spectrometry / methods