Radiolucent lines and osteolysis along tapered straight cementless titanium hip stems: a comparison of 6-year and 10-year follow-up results in 95 patients

Acta Orthop. 2006 Dec;77(6):871-6. doi: 10.1080/17453670610013150.

Abstract

Background: The progress of radiolucent lines (RLs) around the stems of cementless hip endoprostheses and their implications for implant survival are unclear.

Patients and methods: 150 consecutive total hip replacements (THRs) were performed between January and July 1993. The presence and development of RLs and osteolyses were monitored with a standardized special radiographic technique, and the results were compared at 6 and 10 years of follow-up in 95 patients.

Results: At 6 years, RLs were detectable in 43 of 95 patients. RLs with an extension of 50% of the respective zone were present in 35 patients, almost all in Gruen zones 1 and 7. 2 patients presented with osteolysis. At 10 years, the RLs were almost identical--both in number and thickness. Only 1 additional osteolytic lesion had developed; no cup osteolyses were detected, and there was no significant polyethylene wear.

Interpretation: The RLs detected at 6 years were virtually unchanged at 10 years. As they did not progress, they would not jeopardize implant stability. However, the osteolyses seen in 3 patients emphasize the importance of follow-up examinations.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / adverse effects*
  • Female
  • Follow-Up Studies
  • Hip Prosthesis / adverse effects*
  • Humans
  • Joint Instability / etiology
  • Male
  • Middle Aged
  • Osteolysis / diagnostic imaging
  • Osteolysis / etiology*
  • Prosthesis Design
  • Prosthesis Failure
  • Radiography
  • Titanium

Substances

  • Titanium