A solid-phase extraction method for high-performance liquid chromatographic determination of salvianolic acid B in rabbit plasma: application to pharmacokinetic study

Biomed Chromatogr. 2007 Feb;21(2):217-20. doi: 10.1002/bmc.741.

Abstract

A sensitive solid-phase extraction/high-performance liquid chromatographic method with ultraviolet detection was established for the analysis of salvianolic acid B in rabbit plasma. The analyte was separated on a reversed-phase column with trifluoroacetic acid-methanol-acetonitrile (70:10:20, v/v/v) as mobile phase at a flow rate of 1 mL/min, and ultraviolet detection at 315 nm. The calibration curve for salvianolic acid B was linear over the range 35-1400 microg/L with coefficients of correlation >0.999. The inter-day and intra-day precisions of analysis were <15%, and assay accuracy ranged from 95.3 to 109.1%. This method is suitable for determining salvianolic acid B in plasma and thus investigating the pharmacokinetics of salvianolic acid B.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Animals
  • Benzofurans / blood*
  • Benzofurans / pharmacokinetics*
  • Chromatography, High Pressure Liquid / methods*
  • Female
  • Male
  • Rabbits
  • Sensitivity and Specificity
  • Solid Phase Extraction / methods*

Substances

  • Benzofurans
  • salvianolic acid B