Free-flow oxygen delivery to newly born infants

Arch Dis Child Fetal Neonatal Ed. 2007 Mar;92(2):F132-4. doi: 10.1136/adc.2006.103077. Epub 2006 Nov 1.

Abstract

Resuscitation guidelines recommend administration of free-flow oxygen to newly born infants who breathe but remain cyanosed. Self-inflating resuscitation bags are described as unreliable for this purpose. We measured oxygen concentrations >or=80% delivered through a 240 mL Laerdal self-inflating resuscitation bag and from 5 mm tubing inside a cupped hand.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cyanosis / therapy
  • Humans
  • Infant, Newborn
  • Manikins
  • Masks
  • Oxygen Inhalation Therapy / instrumentation*
  • Oxygen Inhalation Therapy / methods
  • Perinatal Care / methods*
  • Resuscitation / instrumentation