Model-based RSA of a femoral hip stem using surface and geometrical shape models

Clin Orthop Relat Res. 2006 Jul:448:92-7. doi: 10.1097/01.blo.0000224010.04551.14.

Abstract

Roentgen stereophotogrammetry (RSA) is a highly accurate three-dimensional measuring technique for assessing micromotion of orthopaedic implants. A drawback is that markers have to be attached to the implant. Model-based techniques have been developed to prevent using special marked implants. We compared two model-based RSA methods with standard marker-based RSA techniques. The first model-based RSA method used surface models, and the second method used elementary geometrical shape (EGS) models. We used a commercially available stem to perform experiments with a phantom as well as reanalysis of patient RSA radiographs. The data from the phantom experiment indicated the accuracy and precision of the elementary geometrical shape model-based RSA method is equal to marker-based RSA. For model-based RSA using surface models, the accuracy is equal to the accuracy of marker-based RSA, but its precision is worse. We found no difference in accuracy and precision between the two model-based RSA techniques in clinical data. For this particular hip stem, EGS model-based RSA is a good alternative for marker-based RSA.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arthroplasty, Replacement, Hip / instrumentation*
  • Femur / surgery*
  • Hip Prosthesis / standards*
  • Humans
  • In Vitro Techniques
  • Models, Theoretical*
  • Photogrammetry
  • Prosthesis Design
  • Reproducibility of Results
  • Surface Properties