Most commercial insulin assays fail to detect recombinant insulin analogues

Ann Clin Biochem. 2006 Jul;43(Pt 4):306-8. doi: 10.1258/000456306777695690.

Abstract

Insulin assays are utilized in various clinical scenarios, including the assessment of insulin therapy compliance or of suspected insulin overdose. In an interpretative exercise carried out by UK National External Quality Assessment Service (NEQAS), serum sent to the participating laboratories was spiked with 30 pmol/L of the short-acting insulin analogue Human Actrapid. Only two out of 24 participant laboratories had sufficient assay cross-reactivity with Actrapid to interpret the results as suggestive of insulin administration. The development of specific insulin assays has led to deterioration in the ability to detect non-compliance or overdose with recombinant insulin treatment. Clinicians should be aware of this significant limitation, which could lead to misdiagnosis.