Multi-residue liquid chromatography/tandem mass spectrometric analysis of beta-agonists in urine using molecular imprinted polymers

Rapid Commun Mass Spectrom. 2005;19(19):2801-8. doi: 10.1002/rcm.2126.

Abstract

Ion suppression, a matrix effect that affects quantitative mass spectrometry, is one of the main problems encountered in liquid chromatography/tandem mass spectrometry. Two different clean-up steps for the multi-residue analysis of beta-agonists in urine were evaluated with respect to minimisation of ion suppression, namely, a mixed-phase solid phase extraction (SPE) column, i.e., clean screen Dau (CSD), and a molecular imprinted polymer (MIP) SPE column. Ion suppression experiments revealed that CSD sample clean-up can lead to false negative results for some beta-agonists, and that clean-up using MIP columns is more selective for beta-agonists than the use of CSD columns.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adrenergic beta-Agonists / urine*
  • Artifacts*
  • Chromatography, High Pressure Liquid / methods*
  • False Negative Reactions
  • Polymers*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Surface Properties

Substances

  • Adrenergic beta-Agonists
  • Polymers