Validation of a 40 MHz B-scan ultrasound biomicroscope for the evaluation of osteoarthritis lesions in an animal model

Osteoarthritis Cartilage. 2005 Feb;13(2):171-9. doi: 10.1016/j.joca.2004.10.022.

Abstract

Objective: To evaluate high frequency (40 MHz) B-mode ultrasound for the detection of osteoarthritis (OA) lesions of varying severity in an animal model of OA.

Design: Ultrasound biomicroscopy (UBM) was performed on the femoral articular surface of adult rabbits with unilateral transection of the anterior cruciate ligament at 4, 8 and 12 weeks post-surgery and on control rabbits. The articular cartilage was examined and graded macroscopically and histologically for OA lesions. Histological examination was used as a reference to determine sensitivity and specificity of ultrasonographic and macroscopic examination regarding fibrillation and ulceration of articular cartilage.

Results: Identification of slight surface irregularities was made possible with UBM. The sensitivity and specificity of UBM were 92.3% and 96.4%, respectively, to detect histological fibrillation and 90.9% and 97.6%, respectively, to identify histological ulceration. Macroscopic examination using India Ink had a sensitivity and specificity of 80% and 96.4%, respectively, for fibrillation and 90.9% and 90.5%, respectively, for ulceration when compared to histology. A high correlation (rsp=0.90) was found between ultrasonographic and histological scores.

Conclusions: UBM of articular cartilage reflects histological structure and can accurately detect early changes such as fibrillation. UBM has the potential to be a valuable tool for the in vivo identification of early lesions of OA and for monitoring the disease or efficacy of novel therapy if it can be packaged in a minimally invasive format suitable for intra-articular imaging.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Anterior Cruciate Ligament / surgery
  • Cartilage, Articular / diagnostic imaging*
  • Cartilage, Articular / pathology
  • Disease Models, Animal
  • Femur
  • Male
  • Microscopy, Acoustic / methods*
  • Osteoarthritis / diagnostic imaging*
  • Osteoarthritis / pathology
  • Rabbits
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Severity of Illness Index