Purpose: To prospectively evaluate the use of a recently developed expanded polytetrafluoroethylene (PTFE)-covered nitinol stent-graft in preventing the need for repeated intervention after transjugular intrahepatic portosystemic shunt (TIPS) creation.
Materials and methods: Fifty-three consecutive patients underwent TIPS procedures between January 2000 and February 2002. Minimum patient follow-up was 9 months (mean, 16.3 months). Fifty-six stent-grafts were implanted in 53 patients; eight of the devices were 8 mm in diameter and 48 were 10 mm in diameter. The stent length varied from 4 to 7 cm. Indications for the procedure included recurrence of bleeding after sclerotherapy (28 patients with cirrhosis, one patient without), refractory ascites or hydrothorax (21 patients with cirrhosis, one patient without), and Budd-Chiari syndrome (two patients).
Results: A technical success rate of 100% was obtained, with an early clinical success rate of 96.2%. During the follow-up period, the recurrence rate was 3.4% (one of 29 patients) for bleeding and 9.0% (two of 22 patients) for ascites. Shunt malfunction occurred in nine of 53 patients (16.9%); in one of these nine patients, shunt occlusion was evident after revision, and a parallel shunt was created. The 1-year primary and secondary patency rates were 83.8% and 98.1%, respectively. In this series, the incidence of encephalopathy (included even as a single short-lived episode) was 47.1% (25 of 53 patients). The 30-day mortality rate was 3.8% (two of 53), and the late mortality rate was 17.3% (eight of 46), excluding seven patients who underwent transplantation.
Conclusion: The new PTFE-covered nitinol stent-graft used appears to be excellent in preventing the need for repeated interventions. A primary patency rate of 83.8% and a secondary patency rate of 98.1% were achieved.
Copyright RSNA, 2004