Validation of coronary flow reserve measurements by thermodilution in clinical practice

Eur Heart J. 2004 Feb;25(3):219-23. doi: 10.1016/j.ehj.2003.11.009.

Abstract

Background: Coronary flow reserve (CFR) and fractional flow reserve (FFR) provide complementary information on the coronary circulation. Using a pressure wire, it is possible to calculate CFR by thermodilution (CFR(thermo)), so that FFR and CFR can be measured with a single guide wire. The present multicentric study was performed to compare the feasibility of CFR(thermo)obtained with an improved algorithm and a standardized injection technique and its agreement with Doppler-derived CFR (CFR(Doppler)).

Methods and results: In 86 patients with coronary artery disease recruited during 1 week in eight centres FFR, CFR(thermo)and CFR(Doppler)were measured. FFR could be obtained in all patients (100%). An optimal CFR(Doppler)could be obtained in 69% of the patients. CFR(thermo)could be obtained in 97% of the patients. A significant correlation was found between CFR(Doppler)and CFR(thermo)(r=0.79, P<0.0001) but CFR(thermo)tended to be higher than CFR(Doppler).

Conclusions: In a setting close to 'real world' practice, this multicentric study confirms the feasibility and reliability of thermodilution-derived CFR. In addition, the safety and the swiftness of assessing FFR and CFR with one single guide wire makes the latter a unique clinical tool for the evaluation of the coronary circulation.

Publication types

  • Comparative Study
  • Multicenter Study
  • Validation Study

MeSH terms

  • Coronary Circulation / physiology*
  • Feasibility Studies
  • Humans
  • Thermodilution / methods
  • Ultrasonography, Doppler