Extracranial stereotactic radiotherapy: evaluation of PTV coverage and dose conformity

Z Med Phys. 2002;12(4):221-9. doi: 10.1016/s0939-3889(15)70476-9.

Abstract

During the past few years the concept of cranial stereotactic radiotherapy has been successfully extended to extracranial tumoral targets. In our department, hypofractionated treatment of tumours in lung, liver, abdomen, and pelvis is performed in the Stereotactic Body Frame (ELEKTA Instrument AB) since 1997. We present the evaluation of 63 consecutively treated targets (22 lung, 21 liver, 20 abdomen/pelvis) in 58 patients with respect to dose coverage of the planning target volume (PTV) as well as conformity of the dose distribution. The mean PTV coverage was found to be 96.3% +/- 2.3% (lung), 95.0% +/- 4.5% (liver), and 92.1% +/- 5.2% (abdomen/pelvis). For the so-called conformation number we obtained values of 0.73 +/- 0.09 (lung), 0.77 +/- 0.10 (liver), and 0.70 +/- 0.08 (abdomen/pelvis). The results show that highly conformal treatment techniques can be applied also in extracranial stereotactic radiotherapy. This is primarily due to the relatively simple geometrical shape of most of the targets. Especially lung and liver targets turned out to be approximately spherically/cylindrically shaped, so that the dose distribution can be easily tailored by rotational fields.