A reversed-phase high performance liquid chromatographic(RP-HPLC) method for the direct determination of three purine bases(theobromin, theophyllin and caffeine) in tea was developed. An ODS column with Zorbax SB-C18(4.6 mm i.d. x 250 mm, 5 microns) was employed. The aqueous solution of methanol containing 0.05% of acetic acid and 0.25% of N,N-dimethylformamide(DMF) was used as eluent with a flow rate of 0.8 mL/min. In this method, the aqueous extract of tea can be injected into HPLC directly, but in current HPLC methods for purine bases the coexisted tea polyphenols must be pre-separated. The three purine bases in tea were separated without any interference from the coexisted tea polyphenols. This method is simple (without any special sample pretreatment) and sensitive with detection limits (S/N = 3) of 0.7, 0.9 and 1.8 mg/L for theobromin, theophyllin and caffeine respectively. The linear range of the calibration curve of peak area for the three purine bases were from 6 mg/L to 1,000 mg/L with a correlation coefficient (r) of 0.998-0.999.