Despite the cyclotron production method and the efficiency of the radiochemical procedures adopted, the long-lived radio-isotopic impurity 202Tl is always present in [201Tl]-labelled radio-pharmaceuticals, together with other short-lived impurities like, 200Tl. Rapid determination of the 202Tl impurity, can be achieved using HPGe gamma spectrometry and a detector shielded by a 5 mm thick envelope of lead. In this way, dead-time correction errors, Compton and X-ray background, are very efficiently avoided and suppressed. The same method could be applied routinely in nuclear medicine, to determine the radioisotopic purity of 201Tl by means of an ionisation chamber dose calibrator.