Rapid HPLC measurement of buflomedil in plasma in poisoning cases

Biomed Chromatogr. 2002 Oct;16(7):482-5. doi: 10.1002/bmc.189.

Abstract

A rapid high-performance liquid chromatographic method for the determination of buflomedil in human plasma is described. It requires a single liquid-liquid extraction step from 1 mL of plasma with diethyl ether followed by chromatography on a Nova Pak C(18) reversed-phase column and detection by ultaviolet light. Metoclopramide was used as internal standard. The method is sensitive with a quantification limit at 500 ng/mL. It was used for the determination of buflomedil in biological fluids in poisoning cases.

MeSH terms

  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Poisoning / blood*
  • Pyrrolidines / blood*
  • Pyrrolidines / poisoning
  • Reference Standards
  • Reproducibility of Results
  • Vasodilator Agents / blood*
  • Vasodilator Agents / poisoning

Substances

  • Pyrrolidines
  • Vasodilator Agents
  • buflomedil