Regulating arsenic species in drinking waters is a reasonable objective, since the various species have different toxicological impacts. However, developing robust and sensitive speciation methods is mandatory prior to any such regulations. Numerous arsenic speciation publications exist, but the question of robustness or ruggedness for a regulatory method has not been fully explored. The present work illustrates the use of anion exchange chromatography coupled to ICP-MS with a commercially available "speciation kit" option. The mobile phase containing 2 mM NaH(2)PO(4) and 0.2 mM EDTA at pH 6 allowed adequate separation of four As species (As(III), As(V), MMAA, DMAA) in less than 10 min. The analytical performance characteristics studied, including method detection limits (lower than 100 ng L(-1) for all the species evaluated), proved the suitability of the method to fulfill the current regulation. Other parameters evaluated such as laboratory fortified blanks, spiked recoveries, and reproducibility over a certain period of time produced adequate results. The samples analyzed were taken from water utilities in different areas of the United States and were provided by the U.S. EPA. The data suggests the speciation setup performs to U.S. EPA specifications but sample treatment and chemistry are also important factors for achieving good recoveries for samples spiked with As(III) as arsenite and As(V) as arsenate.