Comparison of volatile sulfur compound concentrations measured with a sulfide detector vs. gas chromatography

J Dent Res. 2002 Feb;81(2):140-3.

Abstract

The accuracy of the Halimeter, an inexpensive, simple instrument that measures total breath volatile sulfur compounds (VSCs), has not been adequately tested. We compared Halimeter measurements with those obtained with a specific and sensitive gas chromatographic (GC) technique. The Halimeter gave different, biexponential responses to a constant concentration of different VSCs: The relative response rate and sensitivity were hydrogen sulfide > methyl mercaptan > dimethylsulfide. The transient peak VSC concentration of oral samples was reached long before the sulfide detector fully responded. The GC measurement of initial total VSCs in breath samples was 2.7+/-0.48 times greater than the peak concentration of the Halimeter. However, the plateau phase measurement of the Halimeter was 25% greater than that of GC. While GC and Halimeter measurements positively correlated, appreciable differences were observed. In studies where relatively precise VSC measurements are required, GC is the preferable technique.