Comparison of methods for obtaining longitudinal whole-brain magnetization transfer measurements

J Magn Reson Imaging. 2002 Jan;15(1):8-15. doi: 10.1002/jmri.10040.

Abstract

Purpose: To investigate whether the method of applying image masks can alter quantifiable measures determined from whole-brain MTR calculations.

Materials and methods: Thirty-five T1/MT image pairs were obtained from five normal volunteers. For each pair a mask was used to specify the regions to be analyzed. Using these regions, a histogram was used to calculate seven global MTR metrics. This process was performed three ways: 1) using a unique mask for each T1/MT pair, 2) sharing a single mask for each subject and registering all intrasubject images to the image corresponding to their mask, and 3) sharing a single mask for each subject and transforming that mask into alignment with each of their original T1/MT image pairs.

Results: With respect to the first method, the latter two methods caused small but significant differences in several parameters.

Conclusion: The method of applying image masks can affect whole-brain MTR values.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Algorithms
  • Brain / anatomy & histology*
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged