Feasibility of in vivo thyroid 131I monitoring with an imaging plate

Appl Radiat Isot. 2001 Oct;55(4):513-6. doi: 10.1016/s0969-8043(01)00096-3.

Abstract

A new in vivo thyroid 131I monitoring method was devised by using an imaging plate (IP). A thyroid image obtained with a realistic neck-thyroid phantom showed a unique shape characteristic of the thyroid gland. A 131I thyroid imaging allows visual confirmation of thyroid accumulation of 131I. The detection limit of the IP system of 290 Bq was about 1/100 of the screening level of 30 kBq in cases of public emergencies. The IP system is applicable for thyroid 131I monitoring.

MeSH terms

  • Humans
  • Iodine Radioisotopes* / adverse effects
  • Phantoms, Imaging
  • Radiation Monitoring / instrumentation
  • Radionuclide Imaging
  • Thyroid Gland / diagnostic imaging*
  • Thyroid Gland / radiation effects

Substances

  • Iodine Radioisotopes