An ex vivo study on radiofrequency tissue ablation: increased lesion size by using an "expandable-wet" electrode

Eur Radiol. 2001;11(9):1841-7. doi: 10.1007/s003300100891.

Abstract

The present comparative study was conducted to validate a newly developed "expandable-wet" electrode for an increased lesion size of radiofrequency ablation (RFA) on excised beef liver. The expandable-wet electrode, which allows interstitial hypertonic saline infusion through retractable curved needles, was compared with "expanded-dry" and "unexpanded-wet" electrodes for RFA lesion size and other parameters. A total of 120 lesions were created under 50 W (groups A-C) and 90 W (groups A'-C') power control mode for 10 min at each ablation site with the following groups: group A and A' of expanded-dry electrode (needles deployed but saline uninfused); group B and B' of unexpanded-wet electrode (saline infused but needle undeployed); and group C and C' of expanded-wet electrode (needles deployed and saline infused). Together with lower impedance and higher power output, the lesion size in group C (5.3+/-0.4 cm) and C' (6.0+/-1.0 cm) were significantly larger (P<0.01) than that in group A (3.3+/-0.3 cm) and A' (2.0+/-0.2 cm), and group B (3.8+/-1.0 cm) and B' (2.6+/-0.4 cm). The RFA lesion size can be significantly enlarged when the expandable electrode is complemented with interstitial hypertonic saline infusion. This design may improve the efficacy of RF tumor ablation.

MeSH terms

  • Animals
  • Cattle
  • Electrodes
  • Equipment Design
  • Hyperthermia, Induced / instrumentation*
  • In Vitro Techniques
  • Liver / pathology*