Narrow stereotactic beam profile measurements using N-vinylpyrrolidone based polymer gels and magnetic resonance imaging

Phys Med Biol. 2001 Mar;46(3):783-97. doi: 10.1088/0031-9155/46/3/313.

Abstract

In this work, polymer gel-MRI dosimetry (using VIPAR gels), radiographic film and a PinPoint ion chamber were used for profile measurements of 6 MV x-ray stereotactic beams of 5 and 10 mm diameter. The VIPAR gel-MRI method exhibited a linear dose response up to 32 Gy. VIPAR gels were found to resolve the penumbra region quite accurately, provided that the in-plane image resolution of the related T2-map is adequate (< or = 0.53 mm). T2-map slice thickness had no significant effect on beam profile data. VIPAR measurements performed with a spatial resolution of 0.13 mm provided penumbra widths (80%-20% distance) of 1.34 and 1.70 mm for the 5 and 10 mm cones respectively. These widths were found to be significantly smaller than those obtained with the film (2.23 mm for the 5 mm cone, 2.45 mm for the 10 mm cone) and PinPoint (2.25 mm for the 5 mm cone, 2.52 mm for the 10 mm cone) methods. Regarding relative depth dose measurements, good correlation between VIPAR gel and PinPoint data was observed. In conclusion, polymer gel-MRI dosimetry can provide relatively accurate profile data for very small beams used in stereotactic radiosurgery since it can overcome, to some extent, the problems related to the finite size of conventional detectors.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Gels
  • Magnetic Resonance Imaging*
  • Particle Accelerators
  • Phantoms, Imaging*
  • Pyrrolidinones*
  • Radiation Dosage
  • Radiosurgery*
  • Water
  • X-Ray Film

Substances

  • Gels
  • Pyrrolidinones
  • Water
  • N-vinyl-2-pyrrolidinone