Background and aim of the study: Unstented aortic valve substitutes offer many of the theoretical advantages of homografts such as superior hemodynamic performance and enhanced durability, particularly when inserted as a root. Many of these depend on the maintained flexibility of the valve components. Calcification of the aortic wall may adversely affect these phenomena. Electron beam computed tomography has been used to evaluate aortic wall calcification in patients undergoing aortic root replacement in a prospective randomized trial designed to compare the Medtronic Freestyle valve with the homograft valve replacement.
Methods: Patients were followed with electron beam computed tomography scans of the aortic root at six-monthly intervals after surgery. A calcification score (Hounsfield units) and a calcified volume score (mm3) were obtained from each scan using a new modified technique. Results were related to hemodynamic data from echocardiography. The prevalence of calcification was also related to the homograft donor age.
Results: Seventy-six patients (age range: 40-79 years) were randomized to root replacement with either homograft (n = 31) or Freestyle (n = 45) valves. Fifty-three scans of the aortic root were performed postoperatively in 37 patients. No statistical difference between the two groups was found at six and 12 months after surgery. However, after 18 months the calcified volume score was 5903.8+/-2356.8 mm3 in the homograft versus 2725.6+/-1500.5 mm3 in the Freestyle group (p = 0.017). There was a correlation between calcification score, calcified volume score and left ventricular mass (r = 0.323, p = 0.093 and r = 0.350, p = 0.068, respectively) on the one hand, and calcification score, calcified volume score and valve size on the other hand (r = 0.178, p = 0.466 and r = 0.068, p = 0.780, respectively).
Conclusions: Electron beam computed tomography provides a powerful tool for the detection of calcium in the aortic wall of valve grafts. There is a low rate of calcification during the first 18 months in the Medtronic Freestyle valve, and this appears to be lower than that observed in homografts. Longer-term follow up of the aortic root in these patients is required. This is an ongoing study.